EN
DSPL-01
DSPL-01

DSPL-01

Guidewires Fracture Fixture

The guidewire is wound around a cylindrical former, then unwound and examined for fractures.

Product Features

Cylinder: The diameter of the test tool cylinder is 10 times the maximum outer diameter of the guide wire;

Bracket: support both ends of the cylinder;

Fixture: clamp the head end of the guide wire;

Regular guide wire size (Test range): 0.30mm, 0.35mm, 0.46mm, 0.53mm, 0.64mm, 0.69mm, 0.71mm, 0.81mm, 0.89mm, 0.97mm, 1.14mm, 1.19mm, 1.27 mm;

Material: 304 stainless steel.

Reference Standard

  • ISO 11070
  • YY0450.1-2020

Related products