DFT-01 Catheter Flow Rate Tester is suitable for the determination of the flow rate of water through the intravascular catheters and non-intravascular catheters, and also be used to test the flow rate of infusion sets.

Product Features

Independent R&D of hydrostatic pressure test device, which ensures the liquid level of 1000±10mm

 

The flow rate is 525ml/min, which can meet the test requirements

 

Equipped with a constant temperature water tank to ensure that the test water temperature is 23 ℃

 

Equipped with a timer for timed testing

 

The uniquely designed up and down lifting structure can be adjusted in height, which is convenient for testing under different water pressures

 

Equipped with a balance station for easy placement of balances or measuring cylinders

 

Equipped with flow regulating valves for easy flow adjustment.

 

DFT-01

Reference Standard

  • ISO8536-4
  • EN1618
  • ISO 10555-1
  • ISO 20698
  • GB15812.1
  • YY0285.1
  • GB8368
  • YY0817-2010

Product Parameter

Applications

 

Basic Applications

Intravascular catheters

Basic Applications

Catheters other than intravascular catheters

Extended Applications

Infusion set

 

Dimensions:

 

constant temperature water tank: 420 (L) x 3809 (W) x 755mm (H)

Test device: 280 (L) x 420 (W) x 1660 (H)

 

Configuration

 

Standard configuration:Instrument, Hydrostatic head, height adjustment device, constant temperature water tank, PU pipe