You might want to know...
You might want to know...
Connecting Expertise, Creating the Future
Contact us to explore professional testing technologies and jointly develop efficient and reliable quality controlsolutions.
Battery Separator Testing Equipment
TQD-06
Carbon Paper Gas Permeability Tester
TQD-06 Carbon Paper Gas Permeability Tester, conforms to GB/T 20042.7-2024 "Proton exchange membrane fuel cells-Part 7: Test method of carbon paper properties, and is used to test the gas permeability performance in through-plane and in-plane directions.
Share:
Product Features
High-precision electronic airflow and air pressure sensors ensure the accuracy of test data
Embedded high-speed microcomputer chip control
Built-in micro printer with storage function to save more than 50,000 data
10.1-inch HD color LCD screen, real-time display of test data
Imported high-speed and high-precision sampling chips to ensure test accuracy and real-time performance
Key components from imported world-renowned brands, with stable and reliable performance
Overvoltage safety protection design
Energy-saving and environmentally friendly design, ultra-low power consumption
Test Principle
Through-plane direction
Install the test cell on the test device, adjust the throttle valve, use a differential pressure gauge to control a certain pressure difference, stabilize it at room temperature and a certain pressure difference for at least 5 minutes, and calculate the flow rate and pressure difference based on the flow meter reading to obtain the gas permeability in through-plane direction.
In-plane direction
Place the sample between the upper and lower clamps so that the sample surface can be completely covered by the clamps. Apply pressure to the clamps so that the sample is subjected to a pressure of 1 MPA. Adjust the throttle valve and use a differential pressure gauge to control a certain pressure. Keep it stable at room temperature and a certain pressure difference for at least 15 seconds. Calculate the flow and differential pressure based on the flow rate and pressure difference to obtain the gas permeability in in-plane direction.
Reference Standard
Product Parameter
|
Specifications |
TQD-06 |
|
Test Range of Through-plane Pressure Difference |
0 ~ 1000 Pa (customizable) |
|
Test Range of Through-plane Flow |
0 ~ 200mL/min customizable) |
|
Test Range of In-plane Pressure Difference |
0-50Kpa customizable) |
|
Test Range of In-plane Flow |
0 ~ 200mL/min (customizable) |
|
Specimen Size |
Through-plane area:>2cm² |
|
Specimen Size |
In-plane: inner radius 4mm, outer radius 22.5mm |
|
Port Size |
Φ6 mm PU Tubing |
|
Instrument Dimension |
500mm (L)×400 mm (W)×380mm (H) |
|
Power Supply |
220VAC 50Hz / 120VAC 60Hz |
|
Net Weight |
26Kg |
Applications
|
Basic Application |
Carbon Paper |
Test the gas permeability of proton exchange membrane fuel cells |
Configuration
Standard configuration:Instrument, LCD touch screen, Micro printer, Fixtures
Note: The gas supply port of the instrument is Φ6 mm PU Tubing